What the TheraTouch SW1 Is Designed For

The TheraTouch SW1 (model SW1000) is a professional radial pressure wave device built for physical therapy, sports medicine, chiropractic, and rehabilitation clinics. It is not a home-use device or a focused shockwave system. It delivers radial pressure pulses through a handpiece to treat minor muscle aches and pains, temporarily increase local circulation, and activate connective tissue.

Before using the SW1, confirm that you are a trained, qualified healthcare professional and that the device is used only by or on the order of a licensed practitioner. Federal law restricts sale to licensed practitioners.

TheraTouch SW1 Key Specifications

  • TheraTouch SW1 Portable Radial Shockwave Device: The device delivers 60–185 mJ at the applicator, with frequency adjustable from 1–22 Hz. Energy and frequency settings directly affect treatment intensity and depth. Maximum energy decreases above 10 Hz.
    TheraTouch SW1 Portable Radial Shockwave Device
    TheraTouch SW1 Portable Radial Shockwave Device
  • TheraTouch SW1 Portable Radial Shockwave Device: It includes 11 body-region presets, a Manual mode, and a Sweep mode, plus storage for up to 30 custom protocols. These modes allow you to start from a clinical starting point or build a fully custom treatment.
  • TheraTouch SW1 Portable Radial Shockwave Device: The device weighs approximately 4.63 lb with battery and has rechargeable lithium-ion battery (22.2V, 2200 mAh) providing about 20,000 shocks per charge at 120 mJ and 10 Hz. Battery operation supports room-to-room mobility without being tied to a wall outlet.

Setting Up the TheraTouch SW1 for Treatment

Follow these steps to prepare the device for a treatment session. Always review the full user manual before first use.

  1. Charge the battery: Insert the rechargeable lithium-ion module and charge for no more than approximately three hours. The device can operate on battery power for about 20,000 shocks at 120 mJ and 10 Hz.
  2. Select and attach the applicator: Choose from the included applicators: 15 mm Steel Round, 15 mm Compound Round, 20 mm Compound Round, 20 mm Steel Round, 35 mm Steel Round, or 35 mm Compound Round. The clinician should select based on target area, treatment technique, tolerance, and protocol—not simply the largest tip.
  3. Apply coupling gel: Coupling gel is not included with the device; supply it separately. Apply a generous layer to the treatment area and the applicator face to ensure proper energy transfer.
  4. Power on and choose a mode: Use the LCD touchscreen, adjustment knob, and shortcut buttons to navigate. Start with a body-region preset or select Manual mode to set energy, frequency, and shock count manually.
  5. Set treatment parameters: Adjust energy (60–185 mJ or 1–4 bar), frequency (1–22 Hz), and shock count (100–10,000 in 100-shock increments). Note that full 185 mJ is not available at every frequency; maximum energy decreases above 10 Hz.
  6. Deliver the treatment: Place the applicator against the skin with firm, steady contact. Move the handpiece over the treatment area. Stop or adjust if the patient reports excessive pain.
  7. After treatment: Clean and inspect the applicator and handpiece. Do not repeat treatment until prior side effects have diminished. The manual requires a 15-minute break after a block reaching 6,000 shocks.

Understanding the Preset Programs

The SW1 offers 11 body-region presets that serve as starting points—not diagnosis-specific prescriptions. For example, a preset for the lower extremity might suggest 1.7 bar, 8 Hz, and 2,000 shocks, while a gluteal preset could recommend 2.7 bar, 10 Hz, and 2,000 shocks. These are not guarantees of correct treatment for every condition. The clinician must still establish the diagnosis, screen contraindications, and adjust parameters accordingly.

Presets do not authorize treatment directly over the spinal column, lungs, large nerves, major vessels, or other contraindicated structures. Always review the complete user manual.

Radial vs. Focused Shockwave: What the SW1 Is Not

The SW1 is a radial pressure wave device, not a focused shockwave system. Radial energy spreads outward from the applicator, making it suitable for superficial to moderate-depth targets (up to approximately 35 mm according to the manufacturer). Focused shockwave systems concentrate energy at a specific depth. Current comparative research has not established one technology as universally superior across all tendon conditions; device choice should reflect the diagnosis, target depth, clinician training, and evidence for the intended application.

Do not use the SW1 for conditions that require focused energy, such as certain deep tendon pathologies where the intended treatment requires focused rather than radial energy.

Safety and Maintenance Tips

Inspect and clean the intended treatment area before each session. Clean and inspect the applicator and handpiece after every use. The handpiece should be serviced only according to the manufacturer’s procedures or by authorized personnel to avoid unexpected downtime.

The device is not intended for frequent transport between separate facilities and is not suitable for home use. Store it in a clean, dry clinical environment. The device warranty is 12 months; the accessory warranty is three years.

Always review the complete user manual rather than relying on this summary. The presence of an anatomical preset does not establish that every diagnosis in that body region is appropriate for treatment.